Prodigy Autocode El manual del propietario Pagina 43

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 45
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 42
Prodigy Diabetes Care, LLC www.prodigymeter.com
PRODIGY
®
Blood Glucose Meters Owner’s Manual
82
83
Prodigy Diabetes Care, LLC 1.800.243.2636
PRODIGY
®
Blood Glucose Meters Owner’s Manual
< 50 mg/dL to over 400 mg/dL measured using both the
manufacturer’s meter and a clinical blood glucose analyzer for
comparison (reference measurement).
Example 1
If the manufacturer’s reference measurement is 70 mg/dL this
means that the meter measurements are considered accurate if they
fall within ± 15 mg/dL of 70 mg/dL (from of 55 mg/dL to 85 mg/dL).
If meter A produces a measurement of 60 mg/dL and meter B
measures 80 mg/dL, then both are in the acceptable range and meet
the accuracy requirement.
Each manufacturer has met the System Accuracy requirement as
part of its testing to be cleared for marketing. In the System Accuracy,
testing a distribution of blood glucose readings will result at the
various glucose concentration levels for one manufacturer’s system.
This is the same for other manufacturer’s systems. Because of the
range of acceptable readings for the accuracy requirement, identical
readings from two different meters on the same blood sample may
not happen.
Symbols Information
Do not re-use. Single use only.
Consult Operating Instructions.
Keep away from sunlight.
Keep dry.
Temperature limitation.
Use by.
Date of manufacture.
Batch code.
Catalog number.
Serial number.
Control.
LOT
REF
SN
CONTROL
Symbol Referent
Vista de pagina 42
1 2 ... 38 39 40 41 42 43 44 45

Comentarios a estos manuales

Sin comentarios